JCAMP-DX Utility Programmes
Author: R.J. Lancashire Department of Chemistry, UWI,
Mona, Kgn7 JAMAICA, ([email protected])
FTtoPC
The first program, FTtoPC, allows connection, via RS-232, from a
Perkin Elmer 1605 FT IR to a PC and captures the intensity
information in the standard
JCAMP-DX version 4.24 DIFDUP compression format.
FTtoPC is a DOS based programme written in
Borland's Turbo Pascal 5. You will need to configure the
communication parameters on the FTIR using the setup softkeys and
be sure to send the data across in JCAMP-DX format. You can use
either COM1 or COM2 with this version
which was last updated 3rd Dec, 1995 (9264 bytes).
Originally this ran on an old XT at 9600 baud. More recently we
have been using a 486 DX and noticed that some lines had dropped
data. I have now lowered the BAUD rate to 2400, where it seems to
work without data loss. (There was no setting for 4800 on the
FTIR).
HPtoJCMP
We have recently purchased another HP 5890 GC and the latest
version comes with a Headspace analyser and PC Datastation. The
output from the GC can be sent to file using the TABULATE
command,
(e.g. TABULATE X1:X2,filename,X
where X1:X2 is the range - default is all
filename is where output is sent
X is the register containing your data)
This produces columns of numbers, such as:
time counts time counts time counts time counts
0.000 49559 0.003 49565 0.007 49565 0.010 49566
0.013 49558 0.017 49562 0.020 49571 0.023 49573
0.027 49563 0.030 49574 0.033 49563 0.037 49565
0.040 49563 0.043 49567 0.047 49565 0.050 49571
where the numbers are incremented across the page, then down.
I have written a simple
conversion program (size = 44765 bytes, last modified 17th
July, 1997) that can be used to generate JCAMP-DX files from
these files.
Be advised that the TABULATE command introduces rounding errors
such that the X values are NOT evenly spaced (deltaX varies). To
correct for this it is possible to import the file into a
spreadsheet then use the FILL/SERIES command to generate new
columns of X values and then resave the file.
MStoJCMP
A number of HP 5890/6890 GCs with MS detector are in Jamaica and
the Department of Chemistry has recently purchased their
own.
Looking at the software that was shipped with the 5973 Mass
Selective Detector, dated October 1996, I realised that the
syntax for the TABULATE command is different to that for the GC
above (it requires double quotes around the file names) and the
output is sorted differently as well, for example:
m/z abund. m/z abund. m/z abund. m/z abund.
50.10 874 61.10 753 75.05 817 88.05 777
51.10 1751 63.05 721 77.05 3386 89.05 580
52.10 881 65.05 1475 78.05 864 91.05 2520
53.10 2625 66.05 1098 79.05 5174 93.05 1090
Note that the data is now arranged from top to bottom and not
left to right.
Since Mass Spectra can be characterised reasonably well from the
20 most intense peaks, I have written a small utility (size= 46411
bytes) to convert the output file from the TABULATE command to a
JCAMP-DX file containing only the top 20 peaks. The program was
last modified, 17th July, 1997.
Our HP ChemStation software (version G1701AA A.00.00) has an
import and export for JCAMP-DX files under the library settings
menu. The data files created from this however are labelled as
JCAMP-DX version 4.10 and do NOT comply with version 4.24 or the
more recently published version 5.0 specifications.
They cannot be viewed with our viewer unless you edit the
files and add/correct the following header items. While some of
this appears cosmetic, without it our program can NOT display
your information
##TITLE=Library Entry 1 $$OK
##JCAMPDX=Revision 4.10 $$ remove the word revision, change to 5.0
##DATA TYPE=MASS SPECTRUM $$OK
$$add ##DATA CLASS= PEAK TABLE
##SAMPLE DESCRIPTION=QI=82 Derivatives; Metals; Misc. Natural products; Fatty acids and Lipids;
##NAMES= NIST 1992 $$OK
##CAS NAME=Methane $$OK
##MOLFORM=CH4 $$ C H 4
##CAS REGISTRY NO=000074-82-8 $$OK
##MP= -300 $$OK
##BP= -300 $$OK
##MW= 16.031 $$OK
##$RETENTION INDEX=0 $$user defined
##$CONDENSED SPECTRUM=NO $$user defined
##NPOINTS= 7 $$OK
$$the following compulsory items are missing:
$$ ##ORIGIN, ##OWNER, ##XUNITS, ##YUNITS, ##XFACTOR, ##YFACTOR
$$ ##FIRSTX, ##LASTX, ##FIRSTY
##XYDATA=(XY..XY) $$ this should be ##PEAK TABLE= (XY..XY)
2 117 $$ ##XYDATA is ONLY used for continuous fixed increment data types
12 380
13 1069
14 2042
15 8879
16 9999
17 164
$$ add ##END=
Note that I am making these programmes freely available and will
not be held accountable for misuse......
Standard Disclaimer applies etc.
Return to Chemistry, UWI-Mona,
Home Page
Copyright © 2000 by Robert John
Lancashire, all rights reserved.
Created and maintained by Robert Lancashire,
University of the West Indies, Mona Campus, Jamaica.
Comments to author at
Robert Lancashire.
Created December 1995. Last modified 19th February 2000.
URL:
http://wwwchem.uwimona.edu.jm:1104/software/jcputil.html